Technical data

Table Of Contents
Serial Bus Commands 28
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 799
CAN Id specification is set by the :SBUS<N>:CAN:TRIGger:PATTern:ID
and:SBUS<N>:CAN:TRIGger:PATTern:ID:MODE commands.
CAN Data specification is set by the :SBUS<N>:CAN:TRIGger:PATTern:DATA
command.
CAN Data Length Code is set by the
:SBUS<N>:CAN:TRIGger:PATTern:DATA:LENGth command.
<N> An integer, 1-4.
Example To enable the SBUS1 trigger, set the trigger to CAN Data frame matching
the specified Standard ID 0x1f, Data 0x7fffff, and DLC 3.
myScope.WriteString ":TRIGger:MODE SBUS1"
myScope.WriteString ":SBUS1:MODE CAN"
myScope.WriteString ":SBUS1:CAN:TRIGger DATA"
myScope.WriteString ":SBUS1:CAN:TRIGger:PATTern:ID:MODE STANdard"
myScope.WriteString ":SBUS1:CAN:TRIGger:PATTern:ID '0x1f'"
myScope.WriteString ":SBUS1:CAN:TRIGger:PATTern:DATA:LENGth 3"
myScope.WriteString ":SBUS1:CAN:TRIGger:PATTern:DATA '0x7fffff'"
Query :SBUS<N>:CAN:TRIGger?
The :SBUS<N>:CAN:TRIGger? query returns the current CAN trigger on
condition.
Returned Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK
| MESS | MSIG}
See Also ":TRIGger:MODE" on page 916
":SBUS<N>:MODE" on page 792
":SBUS<N>:CAN:TRIGger:PATTern:DATA" on page 800
":SBUS<N>:CAN:TRIGger:PATTern:DATA:LENGth" on page 801
":SBUS<N>:CAN:TRIGger:PATTern:ID" on page 802
":SBUS<N>:CAN:TRIGger:PATTern:ID:MODE" on page 803
":SBUS<N>:CAN:SIGNal:DEFinition" on page 796
":SBUS<N>:CAN:SOURce" on page 797
OVERload Overload - Overload Frame
ACKerror Ack Error - Acknowledge Error
Remote <condition> parameter Front-panel Trigger on: softkey selection
(softkey text - softkey popup text)