Technical data

Table Of Contents
Display Commands 16
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 307
:DISPlay:GRATicule:SETGrat
Commands :DISPlay:GRATicule:SETGrat <DispGratChan>,<number>
The :DISPlay:GRATicule:SETGrat command assigns the corresponding
waveform to a specific grid on the display.
<DispGratChan> Can be:
CHN<N>
DIFF1, DIFF3
COMM2, COMM4
MEM<N> where N is between 1 and 4
FN<N> where N is between 1 and 16 (function)
HIST
D<M> where M is between 0 and 15 (on MSO models with 16 digital
channels)
BUS<Y> where Y is between 1 and 4 (on MSO models)
<number> 1-16, the number of the grid you want to assign the waveform to.
Example This example assigns the histogram to grid 2.
myScope.WriteString ":DISPlay:GRATicule:SETGrat HIST,2"
See Also ":DISPlay:GRATicule" on page 304
":DISPlay:GRATicule:INTensity" on page 305
":DISPlay:GRATicule:NUMBer" on page 306
":DISPlay:GRATicule:SIZE" on page 308