Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 237
:CHANnel<N>:PROBe:PRECprobe:CALibration
Command :CHANnel<N>:PROBe:PRECprobe:CALibration <cal_string>[,<cal_string2]
The :CHANnel<N>:PROBe:PRECprobe:CALibration command specifies the
name of the PrecisionProbe or PrecisionCable calibration to use for the
specified channel and probe.
<N> An integer, 1-4.
<cal_string>[,<c
al_string2>]
A quoted string that is the name of the PrecisionProbe or Precision Cable
calibration. The SMA probe heads can use two independent calibration
files.
Example This example says to use the PrecisionProbe or PrecisionCable calibration
named "2- 8- 2" for channel 1.
myScope.WriteString ":CHANNEL1:PROBE:PRECprobe:CALibration "2-8-2""
Query :CHANnel<N>:PROBe:PRECprobe:CALibration?
The :CHANnel<N>:PROBe:PRECprobe:CALibration? query returns the
currently specified name for the selected channel's PrecisionProbe or
PrecisionCable calibration.
Returned Format
[:CHANnel<N>:PROBe:PRECprobe:CALibration] <cal_string>[,<cal_string2]<NL
>
See Also ":CHANnel<N>:PROBe:ACCAL" on page 218
":CHANnel<N>:PROBe:PRECprobe:MODE" on page 238
":CHANnel<N>:PROBe:PRECprobe:ZSRC" on page 239
":CHANnel<N>:PROBe:PRECprobe:BANDwidth" on page 236