Technical data

Table Of Contents
Calibration Commands 11
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 187
:CALibrate:TEMP?
Query :CALibrate:TEMP? CHANnel<N>
The :CALibrate:TEMP? query returns two delta temperature values in
Celsius:
Between the current temp and the temp of the last time scale
calibration.
Between the current temp and the temp of the last regular user
calibration.
Returned Format
[:CALibrate:TEMP] <ts_delta_temp>,<user_delta_temp><NL>
For example, the string returned could be "- 1,0". A difference in the two
delta values of one degree is not uncommon.