Technical data

Table Of Contents
Acquire Commands 9
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 153
:ACQuire:COMPlete:STATe
Command :ACQuire:COMPlete:STATe {{ON | 1} | {OFF | 0}}
The :ACQuire:COMPlete:STATe command specifies the state of the
:ACQuire:COMPlete mode. This mode is used to make a tradeoff between
how often equivalent time waveforms are measured, and how much new
data is included in the waveform record when a measurement is made.
This command has no effect when the oscilloscope is in real time mode
because the entire record is filled on every trigger. However, in equivalent
time mode, as few as 0 new data points will be placed in the waveform
record as the result of any given trigger event. You set the acquire mode of
the oscilloscope by using the :ACQuire:MODE command.
ON Turns the COMPlete mode on. Then you can specify the completion
percent.
OFF When off, the oscilloscope makes measurements on waveforms after each
acquisition cycle, regardless of how complete they are. The waveform
record is not cleared after each measurement. Instead, previous data
points will be replaced by new samples as they are acquired.
Query
:ACQuire:COMPlete:STATe?
The :ACQuire:COMPlete:STATe? query returns the state of the
:ACQuire:COMPlete:STATe mode.
NOTE
The :ACQuire:COMPlete:STATe command is used only when the oscilloscope is operating in
equivalent time mode and a digitize operation is not being performed. The :DIGitize
command temporarily overrides the setting of this mode and forces it to ON.