Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 1081
:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:LTHReshold
Command :TRIGger:ADVanced:VIOLation:TRANsition:SOURce:LTHReshold {{CHANnel<N>},<
level>}
This command lets you specify the source and low threshold for the
trigger violation transition. The oscilloscope finds a transition violation
trigger by looking for any pulses in your waveform with rising or falling
edges that do not cross two voltage levels in the amount of time you have
specified.
<N> An integer, 1-4.
<level> A real number for the voltage threshold level for the trigger transition
violation, and is used with the high and low threshold transition source
commands.
Query
:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:
LTHReshold? {CHANnel<N>}
The query returns the currently defined transition source for the trigger
transition low threshold violation.
Returned Format
[:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:LTHReshold {CHANnel<N>},]
<level><NL>