Technical data

Table Of Contents
1080 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:HTHReshold
Command :TRIGger:ADVanced:VIOLation:TRANsition:SOURce:HTHReshold {{CHANnel<N>},<
level>}
This command lets you specify the source and high threshold for the
trigger violation transition. The oscilloscope finds a transition violation
trigger by looking for any pulses in your waveform with rising or falling
edges that do not cross two voltage levels in the amount of time you have
specified.
<N> An integer, 1-4.
<level> A real number for the voltage threshold level for the trigger transition
violation, and is used with the high and low threshold transition source
commands.
Query
:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:
HTHReshold? {CHANnel<N>}
The query returns the specified transition source for the trigger transition
high threshold violation.
Returned Format
[:TRIGger:ADVanced:VIOLation:TRANsition:SOURce:HTHReshold {CHANnel<N>},]
<level><NL>