Use Transition Violation Mode to find any edge in your waveform that
violates a rise time or fall time specification. Infiniium Oscilloscopes find a
transition violation trigger by looking for any pulses in your waveform
with rising or falling edges that do not cross two voltage levels in the
amount of time you have specified.
The rise time is measured from the time that your waveform crosses the
low threshold until it crosses the high threshold. The fall time is measured
from the time that the waveform crosses the high threshold until it crosses
the low threshold.
":TRIGger:ADVanced:VIOLation:TRANsition" on page 1078
":TRIGger:ADVanced:VIOLation:TRANsition:SOURce" on page 1079
":TRIGger:ADVanced:VIOLation:TRANsition:SOURce:HTHReshold" on
page 1080
":TRIGger:ADVanced:VIOLation:TRANsition:SOURce:LTHReshold" on
page 1081
":TRIGger:ADVanced:VIOLation:TRANsition:TYPE" on page 1082
Source Use Source to select the channel used for a transition violation trigger.
Low Threshold Use Low Threshold to set the low voltage threshold.
High Threshold Use High Threshold to set the high voltage threshold.
Typ e Use Type to select Rise Time or Fall Time violation.
Trigger On Trigger On parameters include > Time and < Time.
> Time Use > Time to look for transition violations that are longer than the time
specified.
< Time Use < Time to look for transition violations that are less than the time
specified.
Time Use Time to set the amount of time to determine a rise time or fall time
violation.
Available trigger conditioning includes HOLDoff and HYSTeresis (Noise
Reject).
Set the Mode
Before Executing
Commands
Before executing the :TRIGger:ADVanced:VIOLation:TRANsition commands,
set the mode by entering:
:TRIGger:MODE ADVanced and
:TRIGger:ADVanced:MODE VIOLation and
:TRIGger:ADVanced:VIOLation:MODE TRANsition
To query the oscilloscope for the advanced trigger violation mode, enter: