Technical data

Table Of Contents
1070 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:ADVanced:VIOLation:SETup:SHOLd:CSOurce:EDGE
Command :TRIGger:ADVanced:VIOLation:SETup:SHOLd:CSOurce:EDGE {RISing|FALLing}
This command specifies the clock source trigger edge for the clock used
for the trigger setup and hold violation. The clock must pass through the
voltage level you have set before the trigger circuitry looks for a setup and
hold time violation.
Query
:TRIGger:ADVanced:VIOLation:SETup:SHOLd:CSOurce: EDGE?
The query returns the currently defined clock source edge for the trigger
setup and hold violation level for the clock source.
Returned Format [:TRIGger:ADVanced:VIOLation:SETup:SHOLd:CSOurce:EDGE]
{RISing|FALLing}<NL>