Technical data

Table Of Contents
1062 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:ADVanced:VIOLation:SETup:HOLD:CSOurce:LEVel
Command :TRIGger:ADVanced:VIOLation:SETup:HOLD:CSOurce:
LEVel {{CHANnel<N>},<level>}
This command specifies the level for the clock source used for the trigger
hold violation. The clock must pass through the voltage level you have set
before the trigger circuitry looks for a setup or hold time violation.
<N> An integer, 1-4.
<level> A real number for the voltage level for the trigger hold violation clock
waveform, and depends on the type of circuitry logic you are using.
Query
:TRIGger:ADVanced:VIOLation:SETup:HOLD:CSOurce:LEVel? {CHANnel<N>}
The query returns the specified clock source level for the trigger hold
violation.
Returned Format
[:TRIGger:ADVanced:VIOLation:SETup:HOLD:CSOurce:LEVel
{CHANnel<N>},] <level><NL>