Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 1059
:TRIGger:ADVanced:VIOLation:SETup:SETup:DSOurce:LTHReshold
Command :TRIGger:ADVanced:VIOLation:SETup:SETup:DSOurce:LTHReshold {{CHANnel<N>}
,<level>}
This command specifies the data source for the trigger setup violation, and
the low- level data threshold for the selected data source. Data is valid
when it is above the high- level data threshold, and when it is below the
low- level data threshold.
<N> An integer, 1-4.
<level> A real number for the data threshold level for the trigger setup violation,
and is used with the high and low threshold data source commands.
Query
:TRIGger:ADVanced:VIOLation:SETup:SETup:DSOurce: LTHReshold? {CHANnel<N>
}
The query returns the specified data source for the trigger setup violation,
and the low data threshold for the data source.
Returned Format
[:TRIGger:ADVanced:VIOLation:SETup:SETup:DSOurce:LTHReshold
{CHANnel<N>},] <level><NL>