33 Trigger Commands
Hold Time Use HOLD to set the amount of hold time used to test for a violation. The
hold time is the amount of time that the data has to be stable and valid
after a clock edge. The minimum is 1.5 ns; the maximum is 20 ns.
Setup and Hold Use SHOLd (Setup and Hold) to set the amount of setup and hold time
used to test for a violation.
The setup time is the amount of time that the data has to be stable and
valid prior to a clock edge. The hold time is the amount of time that the
data waveform has to be stable and valid after a clock edge.
The setup time plus hold time equals 20 ns maximum. So, if the setup
time is 1.5 ns, the maximum hold time is 18.5 ns.
Available trigger conditioning includes HOLDoff and HYSTeresis (Noise
Reject).
Set the Mode
Before Executing
Commands
Before executing the :TRIGger:ADVanced:VIOLation:SETup commands, set
the mode by entering:
:TRIGger:MODE ADVanced and
:TRIGger:ADVanced:MODE VIOLation and
:TRIGger:ADVanced:VIOLation:MODE SETup and
:TRIGger:ADVanced:VIOLation:SETup:MODE <setup_mode>
Where <setup_mode> includes SETup, HOLD, and SHOLd.
To query the oscilloscope for the advanced trigger violation setup mode,
enter:
:TRIGger:ADVanced:VIOLation:SETup:MODE?