1016 Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference
33 Trigger Commands
:TRIGger:ADVanced:DELay:EDLY:ARM:SOURce
Command :TRIGger:ADVanced:DELay:EDLY:ARM:SOURce {CHANnel<N> | DIGital<M>}
This command sets the Arm On source for arming the trigger circuitry
when the oscilloscope is in the Delay By Event trigger mode.
<N> An integer, 1-4.
<M> An integer, 0-15. The digital channels are only available on the MSO
oscilloscopes or DSO models with the MSO license installed.
Query
:TRIGger:ADVanced:DELay:EDLY:ARM:SOURce?
The query returns the currently defined Arm On source for the Delay By
Event trigger mode.
Returned Format
[:TRIGger:ADVanced:DELay:EDLY:ARM:SOURce] {CHANnel<N>| DIGital<M>}<NL>