Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 1009
:TRIGger:ADVanced:STATe:CLOCk
Command :TRIGger:ADVanced:STATe:CLOCk {CHANnel<N> | DIGital<M> | DONTcare
| <digital_channel>}
This command selects the source for the clock waveform in the State
Trigger Mode.
<N> An integer, 1- 2, for two channel Infiniium Oscilloscope. An integer, 1- 4, for
all other Infiniium Oscilloscope models.
<M> An integer, 0-15. The digital channels are only available on the MSO
oscilloscopes or DSO models with the MSO license installed.
Query
:TRIGger:ADVanced:STATe:CLOCk?
The query returns the currently selected clock source.
Returned Format
[:TRIGger:ADVanced:STATe:CLOCk] {CHANnel<N>|DIGital<M>}<NL>