:TRIGger:ADVanced:PATTern:THReshold:LEVel
Command :TRIGger:ADVanced:PATTern:THReshold:LEVel {CHANnel<N>},<level>
The :TRIGger:ADVanced:PATTern:THReshold:LEVel command specifies the
trigger level on the specified channel for the trigger source. Only one
trigger level is stored in the oscilloscope for each channel. This level
applies to the channel throughout the trigger dialogs (Edge, Glitch, and
Advanced). This level also applies to all the High Threshold (HTHReshold)
values in the Advanced Violation menus.
<N> An integer, 1-4.
<level> A real number for the trigger level on the specified channel, External
Trigger, or Auxiliary Trigger Input.
Query
:TRIGger:ADVanced:PATTern:THReshold:LEVel? {CHANnel<N>}
The query returns the specified channel's trigger level.
Returned Format
[:TRIGger:ADVanced:PATTern:THReshold:LEVel {CHANnel<N>},] <level><NL>