Technical data

Table Of Contents
Trigger Commands 33
Agilent Infiniium 9000 Series Oscilloscopes Programmer's Reference 1003
:TRIGger:ADVanced:PATTern:CONDition
Command :TRIGger:ADVanced:PATTern:CONDition {ENTered | EXITed |
{GT,<time>[,PEXits|TIMeout]} |
{LT,<time>} |
{RANGe,<gt_time>,<lt_time>}}
This command describes the condition applied to the trigger pattern to
actually generate a trigger.
<gt_time> The minimum time (greater than time) for the trigger pattern, from 10 ns
to 9.9999999 s.
<lt_time> The maximum time (less than time) for the trigger pattern, from 15 ns to
10 s.
<time> The time condition, in seconds, for the pattern trigger, from 1.5 ns to 10 s.
When using the GT (Present >) parameter, the PEXits (Pattern Exits) or
the TIMeout parameter controls when the trigger is generated.
Query
:TRIGger:ADVanced:PATTern:CONDition?
The query returns the currently defined trigger condition.
Returned Format
[:TRIGger:ADVanced:PATTern:CONDition] {ENTered|EXITed |
{GT,<time>[,PEXits|TIMeout]} | {LT,<time>} | {RANGe,<gt_time>, <lt_time>
}}<NL>