Technical data

29
Agilent Infiniium 9000 Series Oscilloscopes
Evaluation Guide
4
InfiniiScan Triggering
Capturing infrequent or hard to describe anomalies such as random
glitches requires oscilloscopes with innovative technology.
InfiniiScan is a software-based post-processing technology that
makes it easy to isolate anomalies.
This lab requires an InfiniiScan option license.
Make connections and perform initial setup:
1 To power the demo board, connect a USB cable between it and the
Agilent Infiniium 9000 Series oscilloscope.
You can also apply power by connecting the oscilloscope’s MSO
cable to the demo board’s 40-pin connector.
2 Connect the scope channel 1 probe to demo board CH1 and GND.
3 Make sure all other probes are disconnected from the
oscilloscope.
4 Turn the SELECT switch to position 3 (Burst).
5 Press [Default Setup].
6 Press [AutoScale].
7 Set the time base to 1 µs/div.
Set up the trigger:
8 From the on-screen main menu, choose Trigger> InfiniiScan... (or
press [Menu] in the front panel Trigger controls).
9 In the Trigger Setup dialog, select: