User`s guide
2 Initial Setup
54 InfiniiVision Oscilloscopes User’s Guide
Verifying Basic Oscilloscope Operation
If you have a 6000L Series oscilloscope, you need to start a Web control session, as
described in “Accessing the Web Interface” on page 280.
1 Press the [Save/Recall] key on the front panel, then press the Default Setup softkey
([Save/Recall]&Default Setup). The oscilloscope is now configured to its default
settings (configuration details on page 392).
2 Connect an oscilloscope probe from channel 1 to the Probe Comp terminal on the
front panel.
3 Connect the probe’s ground lead to the ground terminal (next to the Probe Comp
terminal).
4 Press [AutoScale].
5 You should see a waveform on the oscilloscope’s display similar to this:
If you see the waveform, but the square wave is not shaped correctly as shown above,
perform the procedure “Compensating Passive Probes” on page 77.
If you do not see the waveform, ensure your power source is adequate, the oscilloscope
is powered-on, the power cord is firmly inserted into the oscilloscope receptacle, and the
probe is connected securely to the front-panel oscilloscope channel input BNC and to the
Probe Comp terminal.