Interpreting I
2
S Decode
Angled waveforms show an active bus (inside a packet/frame).
Mid-level blue lines show an idle bus.
In the decoded data:
Right channel data values appear in green along with the “R:” characters.
Left channel data values appear in white along with the “L:” characters.
Decoded text is truncated at the end of the associated frame when there is
insufficient space within frame boundaries.
Red dots in the decode line indicate that more data can be displayed. Scroll or expand
the horizontal scale to view the data.
Aliased bus values (undersampled or indeterminate) are drawn in red.
Unknown bus values (undefined or error conditions) are drawn in red.