User`s guide
Triggering 4
InfiniiVision Oscilloscopes User’s Guide 101
The CAN baud rate can be set to:
The default baud rate is 1 Mb/s.
If the desired baud rate is not shown in the table, select User Defined. You can set the
CAN baud rate from 10.0 kb/s to 1.000 Mb/s in increments of 100 b/s. Press the User
Baud softkey and rotate the Entry knob to make your selection.
11 Repeatedly press and release the Smpl Pt softkey to select the point between phase
segments 1 and 2 where the state of the bus is measured. This controls the point
within the bit’s time where the bit value is captured.
You can use the Zoom mode for easier navigation of the decoded data.
10 kb/s 50 kb/s 100 kb/s 500 kb/s
20 kb/s 62.5 kb/s 125 kb/s 800 kb/s
33.3 kb/s 83.3 kb/s 250 kb/s 1 Mb/s
User Defined
NOTE
If the setup does not produce a stable trigger, the CAN signal may be slow enough that
the oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
softkey to set the trigger mode from Auto to Normal.
NOTE
To display CAN serial decode, see page 295.