Technical data
12 Using Segmented Memory Acquisition to Capture More Data
64 Agilent InfiniiVision 7000B Series Oscilloscopes Evaluation Guide
Part B: Using Segmented Memory on Packetized Serial
Bus Signals
1 Connect the demo kit’s 40-pin ribbon cable from the back of the Agilent
InfiniiVision Series oscilloscope to the 40-pin connector on the demo
board.
2 Connect channel 1 probe to the CH1 test point and ground (GND).
3 Connect channel 2 probe to the CH2 test point and ground (GND).
4 Set the rotary switch on the demo board for the “CAN” signal (Mode #8).
5 Press [Default Setup].
6 Set the channel 1 vertical scale to
500 mV/div and the vertical offset to
+1.0 V.
7 Push the Trigger Level knob to
automatically set the trigger level to
50%.
8 Press [Serial]; then, select the CAN
serial decode mode.
9 Press Settings.
10 Press Baud ; then, select 125 kb/s as the baud rate
11 Press [Trigger]; then, select the CAN triggering mode.
12 Set the timebase to 100.0 µs/div and the horizontal delay to +400.0 µs.