Technical data
Using Segmented Memory Acquisition to Capture More Data 12
Agilent InfiniiVision 7000B Series Oscilloscopes Evaluation Guide 63
The oscilloscope should have just captured 150 consecutive waveform
segments with Segment #1 shown on-screen. Let’s now review all of the
waveform segments and determine how long of a time-span the
oscilloscope just captured.
11 Press Current Seg.
12 Review all 150 waveform segments by rotating the selection knob.
13 Rotate the selection knob until Segment #150 is selected and shown on
screen.
Notice that each segment/burst has been time-tagged with a time value
relative to the first captured segment (Segment #1). Segment #150
should have a time-tag of approximately 149 ms. Capturing this much
waveform information at a high sample rate using conventional
acquisition memory would be impossible. Let’s now see how Segmented
Memory can be used on packetized serial bus signals.