Technical data

58 7000 Series Oscilloscopes Service Guide
2 Testing Performance
Horizontal t Accuracy
Generator Setting Test Limits Results
Period 100 µs 99.8 µs to 100.2 µs ________
Period 100 ns 99.8 ns to 100.2 ns ________
Period 5 ns 4.96 ns to 5.04 ns ________
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
Internal trigger
1 division at 25 MHz ________ ________ ________ ________
1 GHz models 0.6 division at 1 GHz ________ ________ ________ _______
500 MHz models 0.6 division at 500 MHz ________ ________ ________ _______
350 MHz models 0.6 division at 350 MHz ________ ________ ________ _______
100 MHz models 0.6 division at 100 MHz ________ ________ ________ _______
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3* Channel 4*
External trigger (7052A, 7032A, 7012A)
± 1 V range:
100 mV at 25 MHz ________ ________ ________ ________
7052A 200 mV at 500 MHz ________ ________ ________ ________
7032A 200 mV at 350 MHz ________ ________ ________ ________
7012A 200 mV at 100 MHz ________ ________ ________ ________
External trigger (7052A, 7032A, 7012A)
± 8 V range:
250 mV at 25 MHz ________ ________ ________ ________
7052A 500 mV at 500 MHz ________ ________ ________ ________
7032A 500 mV at 350 MHz ________ ________ ________ ________
7012A 500 mV at 100 MHz ________ ________ ________ ________
External trigger (7104A, 7054A, 7034A, 7014A)
500 mV at 500 MHz ________ ________ ________ ________
* Where applicable