Technical data

Testing Performance 2
7000 Series Oscilloscopes Service Guide 51
Test External Trigger Sensitivity (2-channel models)
Verify the external trigger sensitivity at these settings:
Trigger range = +/- 1 V
500 MHz (MSO/DSO7052A), 200 mV
pp
350 MHz (MSO/DSO7032A), 200 mV
pp
100 MHz (MSO/DSO7012A), 200 mV
pp
Trigger range = +/- 8 V
500 MHz (MSO/DSO7052A), 500 mV
pp
350 MHz (MSO/DSO7032A), 500 mV
pp
100 MHz (MSO/DSO7012A), 500 mV
pp
7000 service guide.book Page 51 Thursday, October 16, 2008 12:18 PM