Technical data
48 7000 Series Oscilloscopes Service Guide
2 Testing Performance
Test limits for:
• Internal trigger on all models:
• < 10 mV/div: greater of 1 div or 5 mV
pp
>=10 mV/div: 0.6 div
• External trigger on all 2-channel models (MSO/DSO7xx2A):
• Trigger range: 1.0V
DC to 100 MHz: < 100 mV
pp
>100 MHz to max bandwidth: < 200 mV
pp
• Trigger range: 8.0V
DC to 100 MHz: < 250 mV
pp
>100 MHz to max bandwidth: < 500 mV
pp
• External trigger on 4-channel models:
(MSO/DSO7104A/7054A/7034A/7014A)
DC to 500 MHz: < 500 mV
pp
Table 11 Equipment Required to Verify Trigger Sensitivity
Equipment Critical Specifications Recommended Model/Part
Signal
Generator
25 MHz, 100 MHz, 350 MHz,
500 MHz, and 1 GHz sine
waves
Agilent N5181A, E4400B, or
8648A
Power splitter Outputs differ < 0.15 dB Agilent 11667A
Cable BNC, Qty 3 Agilent 10503A
Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780
Feedthrough 50
Ω BNC (f) to BNC (m) Agilent 0960-0301
7000 service guide.book Page 48 Thursday, October 16, 2008 12:18 PM