Technical data
Testing Performance 2
7000 Series Oscilloscopes Service Guide 43
15 The result from step 14 should be between +3.0 dB and -3.0
dB. Record the result in the Performance Test Record (see
page 57).
16 Move the power splitter from the channel 1 to the channel 2
input.
17 Turn off the current channel and turn on the next channel
using the channel keys.
18 Repeat steps 3 through 17 for the remaining channels, setting
the parameters of the channel being tested where
appropriate.
7000 service guide.book Page 43 Thursday, October 16, 2008 12:18 PM