Technical data

Testing Performance 2
7000 Series Oscilloscopes Service Guide 39
1 Connect the equipment (see Figure 5).
a Use the N cable to connect the signal generator to the
input of the power splitter input.
b Connect the power sensor to one output of the power
splitter.
c Use an N-to-BNC adapter to connect the other splitter
output to the channel 1 input.
Table 8 Equipment Required to Verify Bandwidth
Equipment Critical Specifications
Recommended
Model/Part
Signal Generator 100 kHz - 1 GHz at 200 mVrms Agilent N5181A,
E4400B, or 8648A
Power Meter/Sensor 1 MHz - 1 GHz ±3% accuracy Agilent E4418B/8482A
Power Splitter outputs differ by < 0.15 dB Agilent 11667A
Cable Type N (m) 24 inch Agilent 11500B
Adapter Type N (m) to BNC (m) Agilent 1250-0082 or
Pomona 3288 with
Pomona 3533
7000 service guide.book Page 39 Thursday, October 16, 2008 12:18 PM