Technical data

Testing Performance 2
7000 Series Oscilloscopes Service Guide 31
Figure 2 Setting Up Equipment for Digital Channel Threshold Accuracy
Te s t
4 Use a BNC-banana cable to connect the multimeter to the
other side of the BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the
calibrator as shown in Figure 2.
6 On the oscilloscope, press the D15-D0 key, then press the
Thresholds softkey, then press the D7 - D0 softkey repeatedly
until the check mark is next to User.
7 Press the User softkey to the right of the D7 - D0 softkey, then
turn the Entry knob ( ) on the front panel of the
Test
Connector
BNC-Banana
cable
Channels
8 - 15
Channels
0 - 7
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
Probe
Cables
7000A Series Oscilloscope
7000 service guide.book Page 31 Thursday, October 16, 2008 12:18 PM