Technical data
102 7000 Series Oscilloscopes Service Guide
5 Replacing Assemblies
To remove the power supply
1 Disconnect all cables from power supply board.
2 Locate and remove using a TORX T10 driver the four screws
securing the power supply assembly to the rear deck.
Figure 38 Removing the power supply