Technical data

Table Of Contents
Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 737
:MEASure:VSTArt
(see page 798)
Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also "Introduction to :MARKer Commands" on page 308
"Introduction to :MEASure Commands" on page 324
":MARKer:Y1Position" on page 315
":MARKer:Y2Position" on page 316
":MARKer:YDELta" on page 317
":MARKer:X1Y1source" on page 311
":MEASure:SOURce" on page 349
":MEASure:TDELta" on page 727
":MEASure:TSTArt" on page 731
NOTE
The short form of this command, VSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 800). The normal short form, VST, would be the same for
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
NOTE
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y1Position command (see page 315) instead.