Technical data

Table Of Contents
Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 731
:MEASure:TSTArt
(see page 798)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1
cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 308
"Introduction to :MEASure Commands" on page 324
":MARKer:X1Position" on page 310
":MARKer:X2Position" on page 312
":MARKer:XDELta" on page 314
":MEASure:TDELta" on page 727
":MEASure:TSTOp" on page 732
NOTE
The short form of this command, TSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 800). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
NOTE
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 310) instead.