Technical data

Table Of Contents
648 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:WAVeform:SEGMented:COUNt
(see page 798)
Query Syntax
:WAVeform:SEGMented:COUNt?
The :WAVeform:SEGMented:COUNt query returns the number of memory
segments in the acquired data. You can use the
:WAVeform:SEGMented:COUNt? query while segments are being acquired
(although :DIGitize blocks subsequent queries until the full segmented
acquisition is complete).
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands.
Return Format
<count> ::= an integer from 2 to 2000 in NR1 format (count set by
:ACQuire:SEGMented:COUNt).
See Also ":ACQuire:MODE" on page 199
":ACQuire:SEGMented:COUNt" on page 203
":DIGitize" on page 162
":SINGle" on page 188
":RUN" on page 186
"Introduction to :WAVeform Commands" on page 630
Example Code "Example Code" on page 204
NOTE