644 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TIMebase:MODE must be set to MAIN.
:ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution. If
AVERage, :ACQuire:COUNt must be set to 1 in order to return more
than the measurement record or precision analysis record.
MAXimum or RAW will allow up to 8,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.
Query Syntax
:WAVeform:POINts:MODE?
The :WAVeform:POINts:MODE? query returns the current points mode.
Setting the points mode will affect what data is transferred. See the
discussion above.
Return Format
<points_mode><NL>
<points_mode> ::= {NORMal | MAXimum | RAW}
See Also "Introduction to :WAVeform Commands" on page 630
":WAVeform:DATA" on page 638
":ACQuire:POINts" on page 200
":SYSTem:PRECision" on page 462
":WAVeform:VIEW" on page 657
":WAVeform:PREamble" on page 645
":WAVeform:POINts" on page 641
":TIMebase:MODE" on page 469
":ACQuire:TYPE" on page 208
":ACQuire:COUNt" on page 197