Technical data

Table Of Contents
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 577
:TRIGger:LIN:TRIGger
(see page 798)
Command Syntax
:TRIGger:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA}
The :TRIGger:LIN:TRIGger command sets the LIN trigger condition to be:
SYNCbreak — Sync Break.
ID — Frame ID.
Use the :TRIGger:LIN:ID command to specify the frame ID.
DATA Frame ID and Data.
Use the :TRIGger:LIN:ID command to specify the frame ID.
Use the :TRIGger:LIN:PATTern:DATA:LENGth and
:TRIGger:LIN:PATTern:DATA commands to specify the data string length
and value.
Query Syntax
:TRIGger:LIN:TRIGger?
The :TRIGger:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA}
Errors "- 241, Hardware missing" on page 757
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:LIN:ID" on page 567
":TRIGger:LIN:PATTern:DATA:LENGth" on page 570
":TRIGger:LIN:PATTern:DATA" on page 568
":TRIGger:LIN:SIGNal:DEFinition" on page 752
":TRIGger:LIN:SOURce" on page 574
NOTE
The ID and DATA options are available when the automotive CAN and LIN serial decode
option (Option AMS) has been licensed.