Technical data

Table Of Contents
508 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:EBURst:COUNt
(see page 798)
Command Syntax
:TRIGger:EBURst:COUNt <count>
<count> ::= integer in NR1 format
The :TRIGger:EBURst:COUNt command sets the Nth edge at burst counter
resource. The edge counter is used in the trigger stage to determine which
edge in a burst will generate a trigger.
Query Syntax
:TRIGger:EBURst:COUNt?
The :TRIGger:EBURst:COUNt? query returns the current Nth edge of burst
edge counter setting.
Return Format
<count><NL>
<count> ::= integer in NR1 format
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:EBURst:SLOPe" on page 510
":TRIGger:EBURst:IDLE" on page 509