478 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TIMebase:WINDow:SCALe
(see page 798)
Command Syntax
:TIMebase:WINDow:SCALe <scale_value>
<scale_value> ::= scale value in seconds in NR3 format
The :TIMebase:WINDow:SCALe command sets the zoomed (delayed)
window horizontal scale (seconds/division). The main sweep scale
determines the range for this command. The maximum value is one half of
the :TIMebase:SCALe value.
Query Syntax
:TIMebase:WINDow:SCALe?
The :TIMebase:WINDow:SCALe? query returns the current zoomed window
scale setting.
Return Format
<scale_value><NL>
<scale_value> ::= current seconds per division for the zoomed window
See Also "Introduction to :TIMebase Commands" on page 468
":TIMebase:RANGe" on page 471
":TIMebase:POSition" on page 470
":TIMebase:SCALe" on page 474
":TIMebase:WINDow:RANGe" on page 477