Technical data

Table Of Contents
What's New 1
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 43
Command Differences From 54620/54640 Series Oscilloscopes
The main differences between the version 1.00 programming command set
for the InfiniiVision 6000 Series oscilloscopes and the 54620/54640 Series
oscilloscopes are related to:
:HARDcopy and :DISPlay command subsystem changes for USB printers
and the high resolution color display.
New standards supported by the :TRIGger:TV commands.
Support for 113xA Series probes.
New "RAW" :WAVeform:POINts option for retrieving raw acquisition
record data.
Discontinuance of the common commands for macros.
More detailed descriptions of the new, changed, obsolete, and discontinued
commands appear below.
New Commands
Command Description
:ACQuire:RSIGnal (see page 201) Selects the 10 MHz reference signal mode.
:CHANnel<n>:PROBe:ID? (see page 241) Returns the type of probe attached to the
specified oscilloscope channel.
:CHANnel<n>:PROBe:STYPe (see page 243) Sets the channel probe signal type (STYPe) to
differential or single-ended when using the
113xA Series probes, and determines how
offset is applied.
:CHANnel<n>:VERNier (see page 248) Specifies whether the channel's vernier (fine
vertical adjustment) setting is ON (1) or OFF (0).
:DIGital<n>:SIZE (see page 254) Specifies the size of digital channels on the
display.
:EXTernal:PROBe:ID (see page 271) Returns the type of probe attached to the
external trigger input.
:EXTernal:PROBe:STYPe (see page 272) Sets the external trigger probe signal type
(STYPe) to differential or single-ended when
using the 113xA Series probes, and determines
how offset is applied.
:HARDcopy:FILename (see page 720) Sets the output filename for print formats
whose output is a file. Replaces the 5462x/4x
:HARDcopy:DESTination (see page 718)
command.
:HARDcopy:PDRiver (see page 724) Sets the hardcopy printer driver.
:HARDcopy:IGColors (see page 723) Specifies whether graticule colors are inverted.