Technical data

Table Of Contents
392 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MTESt:RMODe:FACTion:SAVE
(see page 798)
Command Syntax
:MTESt:RMODe:FACTion:SAVE <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:SAVE command sets saving on mask failures
on or off.
See ":SAVE Commands" on page 415 for more information on save options.
Query Syntax
:MTESt:RMODe:FACTion:SAVE?
The :MTESt:RMODe:FACTion:SAVE? query returns the current mask failure
save setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 372
":MTESt:RMODe:FACTion:MEASure" on page 390
":MTESt:RMODe:FACTion:PRINt" on page 391
":MTESt:RMODe:FACTion:STOP" on page 393
NOTE
Setting :MTESt:RMODe:FACTion:SAVE ON automatically sets
:MTESt:RMODe:FACTion:PRINt OFF.