:MEASure:VPP
(see page 798)
Command Syntax
:MEASure:VPP [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VPP command installs a screen measurement and starts a
vertical peak-to- peak measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VPP? [<source>]
The :MEASure:VPP? query measures the maximum and minimum vertical
value for the selected source, then calculates the vertical peak- to- peak
value and returns that value. The peak-to- peak value (Vpp) is calculated
with the following formula:
Vpp = Vmax - Vmin
Vmax and Vmin are the vertical maximum and minimum values present on
the selected source.
Return Format
<value><NL>
<value> ::= vertical peak to peak value in NR3 format
See Also "Introduction to :MEASure Commands" on page 324
":MEASure:SOURce" on page 349
":MEASure:VMAX" on page 361
":MEASure:VMIN" on page 362
":MEASure:VAMPlitude" on page 358