Technical data

Table Of Contents
What's New 1
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 35
What's New in Version 4.10
New features in version 4.10 of the InfiniiVision 6000 Series oscilloscope
software are:
The ability to trigger on and decode FlexRay serial bus data using a
Decomsys BusDoctor 2 protocol analyzer with a four- channel
mixed- signal oscilloscope that includes the Option FRS license.
The square root waveform math function.
Several new hardcopy printer drivers.
More detailed descriptions of the new and changed commands appear
below.
New Commands
Command Description
:SBUS:BUSDoctor:ADDRess Sets/queries the four fields in the BusDoctor
LAN IP Address.
:SBUS:BUSDoctor:BAUDrate Sets/queries the baud rate for the BusDosctor
from 2.5 Mb/s to 10 Mb/s.
:SBUS:BUSDoctor:CHANnel Sets/queries the FlexRay channel that the
BusDoctor analyzes/preprocesses.
:SBUS:BUSDoctor:MODE Sets/queries the operating mode of the
BusDoctor.
:SBUS:FLEXray:COUNt:NULL? (see page 440) Returns the FlexRay null frame count.
:SBUS:FLEXray:COUNt:RESet (see page 441) Resets the FlexRay frame counters.
:SBUS:FLEXray:COUNt:SYNC? (see page 442) Returns the FlexRay sync frame count.
:SBUS:FLEXray:COUNt:TOTal? (see page 443) Returns the FlexRay total frame count.
:TRIGger:FLEXray:ERRor:TYPE (see page 521) Sets/queries the FlexRay error type to trigger
on.
:TRIGger:FLEXray:FRAMe:CCBase (see
page 523)
Sets/queries the base of the FlexRay cycle
count (in the frame header) to trigger on.
:TRIGger:FLEXray:FRAMe:CCRepetition (see
page 524)
Sets/queries the repetition number of the
FlexRay cycle count (in the frame header) to
trigger on.
:TRIGger:FLEXray:FRAMe:ID (see page 525) Sets/queries the FlexRay frame ID to trigger
on.
:TRIGger:FLEXray:FRAMe:TYPE (see page 526) Sets/queries the FlexRay frame type to trigger
on.
:TRIGger:FLEXray:TIME:CBASe Sets/queries the base of the FlexRay cycle to
trigger on.