Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 289
:FUNCtion:SOURce1
(see page 798)
Command Syntax
:FUNCtion:SOURce1 <value>
<value> ::= {CHANnel<n> | GOFT}
<n>::={1|2|3|4}for4chmodels
<n> ::= {1 | 2} for 2ch models
The :FUNCtion:SOURce1 command is used for any :FUNCtion:OPERation
selection (including the ADD, SUBTract, or MULTiply channel math
operations and the FFT, INTegrate, DIFFerentiate, or SQRT transforms).
This command selects the first source for channel math operations or the
single source for the transforms.
The GOFT parameter is only available for the FFT, INTegrate,
DIFFerentiate, or SQRT functions. It lets you specify, as the function input
source, the addition, subtraction, or multiplication of two channels. When
GOFT is used, the g(t) source is specified by the
:FUNCtion:GOFT:OPERation, :FUNCtion:GOFT:SOURce1, and
:FUNCtion:GOFT:SOURce2 commands.
Query Syntax
:FUNCtion:SOURce1?
The :FUNCtion:SOURce1? query returns the current source1 for function
operations.
Return Format
<value><NL>
<value> ::= {CHAN<n> | GOFT}
<n>::={1|2|3|4}for4chmodels
<n> ::= {1 | 2} for 2ch models
See Also "Introduction to :FUNCtion Commands" on page 278
":FUNCtion:OPERation" on page 285
":FUNCtion:GOFT:OPERation" on page 281
":FUNCtion:GOFT:SOURce1" on page 282
":FUNCtion:GOFT:SOURce2" on page 283
NOTE
Another shorthand notation for SOURce1 in this command/query (besides SOUR1) is
SOUR.