Technical data

Table Of Contents
Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 269
:EXTernal:IMPedance
(see page 798)
Command Syntax
:EXTernal:IMPedance <value>
<value> ::= {ONEMeg | FIFTy}
The :EXTernal:IMPedance command selects the input impedance setting for
the external trigger. The legal values for this command are ONEMeg
(1 M
Ω) and FIFTy (50Ω).
Query Syntax
:EXTernal:IMPedance?
The :EXTernal:IMPedance? query returns the current input impedance
setting for the external trigger.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
See Also "Introduction to :EXTernal Trigger Commands" on page 266
"Introduction to :TRIGger Commands" on page 479
":CHANnel<n>:IMPedance" on page 235
NOTE
You can set external trigger input impedance to FIFTy (50Ω) on the 2-channel, 300 MHz,
500 MHz, and 1 GHz bandwidth oscilloscope models.