Technical data

Table Of Contents
252 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:DIGital<n>:LABel
(see page 798)
Command Syntax
:DIGital<n>:LABel <string>
<string> ::= any series of 10 or less characters as quoted ASCII string.
<n> ::= An integer, 0,..,15, is attached as a suffix to the command and
defines the logic channel that is affected by the command.
The :DIGital<n>:LABel command sets the channel label to the string that
follows. Setting a label for a channel also adds the name to the label list
in non- volatile memory (replacing the oldest label in the list).
Query Syntax
:DIGital<n>:LABel?
The :DIGital<n>:LABel? query returns the name of the specified channel.
Return Format
<label string><NL>
<label string> ::= any series of 10 or less characters as a quoted
ASCII string.
See Also "Introduction to :DIGital<n> Commands" on page 249
":CHANnel<n>:LABel" on page 237
":DISPlay:LABList" on page 262
":BUS<n>:LABel" on page 217
NOTE
This command is only valid for the MSO models.
NOTE
Label strings are 10 characters or less, and may contain any commonly used ASCII
characters. Labels with more than 10 characters are truncated to 10 characters.