Technical data

Table Of Contents
204 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:ACQuire:SEGMented:INDex
(see page 798)
Command Syntax
:ACQuire:SEGMented:INDex <index>
<index> ::= an integer from 2 to 2000 (w/8M memory) in NR1 format
The :ACQuire:SEGMented:INDex command sets the index into the memory
segments that have been acquired.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command. The number of segments to acquire is set using
the :ACQuire:SEGMented:COUNt command, and data is acquired using the
:DIGitize, :SINGle, or :RUN commands. The number of memory segments
that have been acquired is returned by the :WAVeform:SEGMented:COUNt?
query. The time tag of the currently indexed memory segment is returned
by the :WAVeform:SEGMented:TTAG? query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 1M memory allows
a maximum of 250 segments.
Query Syntax
:ACQuire:SEGMented:INDex?
The :ACQuire:SEGMented:INDex? query returns the current segmented
memory index setting.
Return Format
<index><NL>
<index> ::= an integer from 2 to 2000 (w/8M memory) in NR1 format
See Also ":ACQuire:MODE" on page 199
":ACQuire:SEGMented:COUNt" on page 203
":DIGitize" on page 162
":SINGle" on page 188
":RUN" on page 186
":WAVeform:SEGMented:COUNt" on page 648
":WAVeform:SEGMented:TTAG" on page 649
":ACQuire:SEGMented:ANALyze" on page 202
"Introduction to :ACQuire Commands" on page 193
Example Code
' Segmented memory commands example.
' -------------------------------------------------------------------
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.