Product data

Performance characteristics (continued)
Trigger system (continued)
Autoscale Finds and displays all active scope and logic (for 7000 Series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
Scope channel triggering
Range (internal) ±6 div from center screen
Sensitivity* <10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div
Coupling AC, ~10 Hz on MSO/DSO703xA/705xA/710xA),
DC, noise reject, HF reject and LF reject (~50 kHz)
Digital (D15 - D0) channel triggering (7000 Series MSO or MSO-upgraded 7000 Series DSO)
Threshold range (user defined) ±8.0 V in 10 mV increments
Threshold accuracy* ±(100 mV + 3% of threshold setting)
Predefined thresholds TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V
External (EXT) triggering MSO/DSO7xx2A (2-/2+16-ch models) MSO/DSO7xx4A (4-/4+16-ch models)
Input impedance 1 MΩ ± 3% || 14 pF or 50 Ω 2.14 kΩ ±5%
Maximum input CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk ±15 V
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-Ω input
Range DC coupling: trigger level ±1 V and ±8 V ±5 V
Sensitivity For ±1 V range setting: DC to 500 MHz: 500 mV
DC to 100 MHz, 100 mV;
For ±8 V range setting:
DC to 100 MHz, 250 mV;
>100 MHz to bandwidth of oscilloscope: 500 mV
Coupling AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
Probe ID Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.