bus is measured. This controls the point within the bit’s time
where the bit value is captured.
8 Set the trigger level.
• For analog channels, turn the Trigger Level knob.
• For digital channels, press the D15 Thru D0 key and the
Thresholds softkey to access the threshold level setting
softkeys.
9 Return to the previous menu by pressing the up-arrow
softkey.
10 If the decode line does not appear on the display, press the
Decode softkey to turn it on.
11 If the oscilloscope is stopped, press the Run/Stop key to
acquire and decode data.
You can use the Zoom horizontal sweep mode for easier
navigation of the decoded data.
One Bit
60%
70%
80%
Sample Point
NOTE
Changing the Source settings in the Serial Decode setup also changes
them in the Trigger setup.
NOTE
If the setup does not produce a stable trigger, the CAN signal may be slow
enough that the oscilloscope is AutoTriggering. Press the Mode/Coupling
key, then press the Mode softkey to set the trigger mode from Auto to
Normal.