Oscilloscopes User's Guide
Triggering the Oscilloscope 4
6000 Series Oscilloscope User’s Guide 159
In Asynchronous Mode
1 Press the Baud softkey to select a baud rate of 2.5 Mb/s,
5Mb/s, or 10 Mb/s.
2 Press the Channel softkey to select channel A or B.
In Synchronous Mode
1 Press the FIBEX File softkey to specify the location of the
FIBEX (Field Bus Exchange format) file that defines the
time-triggered timing schedule of the probed FlexRay bus.
2 Press the Channel softkey to select channel A or B.
In Asynchronous or Synchronous Mode In addition to the logic
probe cable connection between the VPT1000 and the
oscilloscope, you must make a LAN connection to allow the
oscilloscope to control the VPT1000. See also “Specifying the
VPT1000 LAN Address” on page 157.
NOTE
If the VPT1000 loses synchronization, then press the Mode key (in the
VPT1000 Menu) on the oscilloscope to cycle through the modes and
restore synchronization.
NOTE
Before powering-off the oscilloscope, change the VPT1000 2 mode to PC.
If oscilloscope power is cycled when in Asynchronous or Synchronous
modes, communication with the VPT1000 will be lost. To restore
communication with the VPT1000, cycle the VPT1000 power, wait for its
State indicator to turn from red to green, and then press the Mode key (in
the VPT1000 Menu) on the oscilloscope to cycle through the modes.