88 5000 Series Oscilloscope User’s Guide
2 Front-Panel Controls
To restore the oscilloscope to its default configuration
• Press the Save/Recall key, then press the Default Setup softkey.
The default configuration returns the oscilloscope to its default
settings. This places the oscilloscope in a known operating
condition. The major default settings are:
Horizontal main mode, 100 µs/div scale, 0 s delay (horizontal
position), center time reference.
Vertical Channel 1 on, 5 V/div scale, DC coupling, 0 V position,
1MΩ impedance, probe factor to 1.0 if an AutoProbe probe is
not connected to the channel.
Trigg er Edge trigger, Auto sweep mode, 0 V level, channel 1
source, DC coupling, rising edge slope, 60 ns holdoff time.
Display Vectors on, 33% grid intensity, infinite persistence off.
Other Acquire mode normal, Run/Stop to Run, cursors and
measurements off.
Labels All custom labels that you have created in the Label
Library are preserved (not erased), but all channel labels will be
set to their original names.