User`s guide

Making Measurements 4
5000 Series Oscilloscope User’s Guide 175
Binary — Binary logic levels are displayed directly above
the softkeys for the current X1 and X2 cursor positions for
all displayed channels.
Hex — Hexadecimal logic levels are displayed directly above
the softkeys for the current X1 and X2 cursor positions for
all displayed channels.
In hexadecimal and binary mode, a level can display as 1
(higher than trigger level), 0 (lower than trigger level),
indeterminate state ( ), or X (don't care). In binary mode, X
is displayed if the channel is turned off. In hex mode, the
channel is interpreted as a 0 if turned off.
3 Press the Source softkey to select the oscilloscope channel or
math source on which the Y cursors will indicate
measurements.
The source in Normal cursor mode can be any oscilloscope
channel or math source. If you select binary or hexadecimal
mode, the Source softkey is disabled since you are displaying
binary or hexadecimal levels for all channels.
Cursor X1 values
for channels 1
through 4
Indicates left bit is
channel 1, right bit is
channel 4