User`s guide
Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 105
• 2 ns to 10 s for > or < qualifier (5 ns to 10 s for 100 MHz
and 300 MHz bandwidth models)
• 10 ns to 10 s for >< qualifier, with minimum difference of
5 ns between upper and lower settings
< qualifier time set softkey
• When the less than (<) qualifier is selected, the Entry knob
sets the oscilloscope to trigger on a pulse width less than the
time value displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry
knob sets the upper time range value.
> qualifier time set softkey
• When the greater than (>) qualifier is selected, the Entry
knob sets the oscilloscope to trigger on a pulse width greater
than the time value displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry
knob sets the lower time range value.